Using the LTI-HID USB Interface

The LTI-HID USB Interface provides the same functionality as the current LTI/USB interface, with the addition of the green and red LEDs at the top, which identify what action the interface is currently performing. This is particularly useful if you use multiple interfaces to configure more than one logger at a time.

When you start one of the LogTag® software packages, such as LogTag® Analyzer, LogTag® Connect or LogTag® Xpress, the green LED will turn on to indicate the interface has established connection with the software and is ready for a logger to be inserted.

Note: The green LED will only turn on if the software's Automatic Download feature is enabled. In LogTag® Xpress and LogTag® Connect this is always enabled, in LogTag® Analyzer this feature is enabled by default, but can be turned off in the option settings.

As soon as you insert a logger the automatic download starts, the green LED will turn off and the red LED will turn on. This indicates the software is communicating with the logger, in this case downloading data.

If the communication was successful, the green LED blinks. You can now continue with additional tasks, such as configuring the logger or hibernating it.

When the logger is being configured, either automatically or based on your settings, the red LED also turns on. Once the configuration parameters have been successfully uploaded to the logger, the green LED blinks.

If there was an error during communication, and the data could not be downloaded or uploaded, the red LED blinks. The most common causes for this are a low battery, dirty contacts on the logger or trying to configure a single use logger for a second trip. You will also see the red LED blinking if you don't provide a password where required or cancelling a operation mid way.

 

Table 1: USB Interface LED signals

Action

Green 'Ready' LED

Red 'Busy' LED

Visual

Initialised

simultaneous rapid blinking, 7 times

Not initialised

off

slow blinking

      

No LogTag® software application running

off

off

 

LogTag® software application running, waiting for a logger to be inserted

solid on

off

Logger being downloaded

off

solid on

Download successfully completed, additional communication possible

fast blinking

off

Problems occurred during download (error message, battery low, etc...)

off

fast blinking

Logger being configured

off

solid on

Configuration complete, additional communication possible

fast blinking

off

Problems occurred during configuration (error message, battery low, etc...)

off

fast blinking

Logger being hibernated

off

solid on

Hibernation complete, additional communication possible

fast blinking

off